ThermoBrite® Elite

AUTOMATED LABORATORY ASSISTANT

AUTOMATED SAMPLE PREPARATION WITH KREATECH™ FISH PROBES

The ThermoBrite Elite provides total automation for the pre- and post-hybridization steps in FISH testing  including on-board denaturation and hybridization.

Perfect Solution for Pathology

The pre-installed validated KREATECH™ FISH protocols allow for easy selection of the correct program for your solid tumor, bone marrow and blood samples. Just load your slides and walk away. Minimal hands-on time frees time for other important projects. Adding the probe and placing the coverslip are the only manual steps. After denaturation and hybridization, the system will complete the post- hybridization steps. Just add the DAPI / antifade and coverslip and you are ready to image your slides.

The  ThermoBrite Elite can process up to 12 slides per run. You can also denature and hybridize slides on your ThermoBrite Denaturation and Hybridization System, in order to have the ThermoBrite Elite system available for additional runs.

Interactive easy-to-use Software

The intuitive  easy-to-use software allows you to run standard validated KREATECH™ protocols for blood, bone marrow and solid tumor FFPE samples. The system allows input of 10 different reagents. It has 3 independent waste ports. You can also control the functions in each of the 3 chambers (4 slides per chamber) allowing for small batches or running of 12 slides at one time.

Features

  • Automated fluidic system
  • Accurate temperature control to ±1°C
  • Controlled agitation and mixing
  • User-friendly Graphical User Interface
  • Validated KREATECH™ protocols pre-installed
  • Small benchtop unit

Key Benefits

  • Improves consistency and reproducibility
  • Reduces hands-on time
  • Flexible and easy-to-use
  • Saves time and money

Applications in FISH

  • Pathology (Solid tumor/ FFPE)
  • Hematology (Blood/bone marrow)
  • Cytology (Fluids)
User Interface Steps
 

System Specifications

Slide-based Applications

Histology

Cytology

Hematology

Cytogenetics

Instruments Fluidics

Input Fluidics

Ten reagents input

Output Fluidics

Three waste outputs

Compatibility

Fluid paths compatible w/ common reagents

Dimensions

24"w x 18"d x 15"h, 63 lbs.

Temperature

Control Range

25°C - 95°C

Ramp Rate

5-10°C/min

Ambient

10°C - 37°C, 10%-90% relative humidity

User Interface and Control

PC-based GUI

Continuous data logging

Safety & Regulations

ETL, CSA, CE (pending)

Power Requirements

100-240 VAC, 50/60 Hz, 8A

Intuitive user interface simplifies building or modifying protocols